Configuring your Abaxis HMII / HM2 / HM5 with DVMAX
Configuring your Abaxis HMII / HM2 / HM5 with DVMAX
The following document outlines how to configure your Abaxis Hematology Unit with DVMAX.
Note: Abaxis HMII/HM2 Connection Settings (HMII Only Connects via Serial Connection). HM2/HM5 connects with Serial or USB ports.
- Click on the Bolt Icon
- Click #5 Settings
- Click #2 Customize
- Click #1 General Settings:
- Change PC Link to 9600 (if using null modem serial cable) or USB if using USB cable,
- Change Serial Protocol Version to 3.10
- Change VSx Link to Offline.
- Click Accept